Operating Manual V1.1 – 05/2007
3. Instrument Characteristics
LED for feed rate
LED on, for enabled
amplitude adjustment
Sets the current specimen
position between 0 (lower)
and 20,000μm (upper) or –
after zero, totals the
section thicknesses
Toggles between SPEED
(blade feed rate) and
AMPL (amplitude)
Pressing this button toggles
between μm and μm
(actively illuminated)
Moves the blade
towards the specimen
Moves the blade away
from the specimen
Quickly moves the speci-
men downwards (press
once to adjust to the lowest
position)
Immediately starts or stops
the sectioning process
Setting dial 1, for blade
feed rate and amplitude
Setting dial 2, for
section thickness and
moving the specimen
vertically
Selects the step size
in 1μm,1m or
100μm
increments
Fig. 10
Resets the section
thickness addition
display μm to zero
Press this button
(for approx. 3 sec.) to save a
frequently used feed value
(pressing it more than once
feeds by the saved section
thickness multiple times)
Section thickness
LED display for
section thickness and
section thickness addition
LED display for
blade feed rate and
amplitude
ON/OFF lamp
3.3 VT1200 Control Panel
The Leica VT1200 is a semiautomatic microtome with a vibrating blade. Before each cut,
a manual feed to the desired section thickness must be carried out using the setting dial for
section thickness. The VT1200 does not include an automatic specimen retraction; however,
retraction can be performed manually.
Quickly moves the speci-
men upwards (press the
button until the desired
position is reached)
Caution: Press the button until the desired
position is reached. When the end position is
reached, the LED lights up.