18
Operating Manual V1.1 – 05/2007
5. Working with the Instrument
5.1 Description of the Typical Application
The Leica VT1200 / VT1200S is a microtome with a vibrating blade and is
predominately used for sectioning fixed and unfixed specimens during neu-
rological research.
To simplify the process of inserting the spec-
imen, you can quickly move the specimen
receptacle to the lowest position by pressing
the DOWN button.
To prepare high-quality sections, especially
for unfixed tissue, Leica suggests using the
optional VibroCheck measuring instrument to
determine the height amplitude of the blade
after each blade change, and then to minimize
it using the setting screw on the blade holder.
To perform the aforementioned process, in-
stall the VibroCheck instrument (see p. 36 for
the VT1200 or p. 38 for the VT1200S), install
the blade and adjust the clearance angle to
the desired position. Then take the measure-
ment and make the appropriate adjustments
to the placement of the blade holder.
Remove the VibroCheck according to the in-
structions, and turn the blade 90° toward the
top to ensure that the ice bath and buffer tray
can be installed safely.
Insert the buffer tray into the ice bath and
cover it with the lid. Fill the ice bath with
crushed ice.
Remove the cover and fill the buffer tray with
pre-cooled physiological buffer.
Push the ice bath and buffer tray onto the
dovetail guide and clamp them down.
Use cyanoacrylate adhesive to adhere the
specimen to the specimen plate and use a
manipulator to place it in the buffer tray.
Insert the hose for gassing the buffer into the
hose clamp.
Use the UP button to lift the specimen to
the sectioning level of the blade quickly. Fine-
adjusting the blade-to-specimen proximity can
be performed using the desired step size 1, 10
or 100μm.
Advance the blade toward the specimen us-
ing the "Blade forward" key.
After sectioning the specimen, use the setting
dial to feed to the desired section thickness
using the selected step size (1, 10 or 100μm).
This raises the specimen receptacle to the
desired value.
Start the sectioning process by pressing the
RUN/STOP key. The process can be stopped
again by pressing this key after the cut has
been completed. Use the "Blade back" key
to move the blade in front of the specimen.
Select the section thickness for the next cut
and start the sectioning process again.
After completing the sectioning process,
remove the blade, dispose of the specimen
receptacle by pushing the DOWN key into its
lowest position, and unclamp, empty and clean
out the ice bath and buffer tray.