Leica VT1200 and VT1200 S – Microtome
5.2 Control Elements on the VT1200 Control Panel
Caution: To practice the key functions, always make sure there are
no blades installed! Do not install the blades until you are familiar
with all of the key functions.
When the instrument is switched on, to make it easier to insert a specimen,
the specimen receptacle automatically moves into the lowest position
(the LED in the DOWN key lights up) and the blade holder moves into the
rearmost position (the LED in "Blade away" from the specimen lights up).
The last parameters used before the instrument was switched off are called
up again: blade feed rate (SPEED), selected amplitude (AMPL) and the sec-
tion thickness that was saved using the MEMO key.
The LED illumination is switched on automatically. You can switch it off using
the ON/OFF key (lamp symbol).
5.2.1 Moving the Specimen Receptacle Vertically
When the DOWN key is pressed, the specimen receptacle automatically
moves quickly into the lowest position. (The LED in the DOWN key lights up
when the lower end position is reached.) The μm display is set to "0". If the
DOWN key is pressed a second time while the specimen receptacle is mov-
ing downwards, the specimen receptacle comes to a stop. The current
position appears on the μm display (lowest position = 0). The μm display
does not change while the specimen receptacle is in motion.
When the UP key is pressed and held, the specimen receptacle moves quickly
upwards into the desired position. After the UP key has been released, the
current position of the specimen receptacle appears on the μm display.
If the specimen receptacle reaches the upper end position, the LED in the
UP key lights up (uppermost position = 20,000μm). The μm display does
not change while the specimen receptacle is in motion.
You can reset the display for the current position of the specimen holder to
zero at any point using the CLEAR/μm key. The section thicknesses are
then added in the μm display. Switch to μm by pressing the μm/
μm key.
5. Working with the Instrument
Switch on the
instrument.
1 = on 0 = off