Leica VT1200S Manuel d'utilisateur Page 31

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 58
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 30
31
Leica VT1200 and VT1200 S – Microtome
5. Working with the Instrument
semiautomatic sectioning mode
Rotary knob 2
The feed rate (FOR/REV) for the
"Blade forward" and the "Blade
back" keys can be set between 1
and 5 mm/s in 0.5 mm/s increments
with rotary knob 1.
Rotary knob 2
Switches the vibration movement
(FOR/VIB) of the "Blade forward"
key either on or off using rotary
knob 1.
Rotary knob 2
Key acknowledgment (BEEP) on or
off with rotary knob 1.
Save the parameter and leave by
pressing the menu MENU key.
Leaving the menu and
saving the parameter
is possible at any point
in the menu.
Key/setting dial automatic sectioning mode
ditto
ditto
ditto
ditto
1
2
1
2
1
2
If you want to call up the parameters of a certain user (for example, user 3), press the Menu
key and select user 3. Then confirm by pressing the Menu key 2x. The parameters stored under
user 3 are now enabled.